Swema FLOW126 Air Flow Hood

Model: FLOW126
Brand: Others

Product Details

SwemaFlow 126 measures both exhaust and supply air. For supply air the longer 650x650 mm and 250x650 mm hoods are recommended. The range is air flow from 1.5 to 125 l/s, 5.4 to 450 m³/h, 2.3 to 263 cfm, temperature 0 to 50 °C, 32 to 122 °F, and barometric pressure 600 to 1200 hPa, 18 to 35 inHg. SwemaFlow 126 is based on the well known Swema principle, a net of hot wires, which gives a good and accurate mean value of the air flow in the capture opening.
This principle together with the longer hoods (650x650 or 650x250) makes it also possible to cope with the airflow from angled outlets. Complement the 650x650 hood with a cross to disrupt the swirl created by swirl diffusers. The included 300x300x470 red foldable hood could measure supply air after calibrating the result with a longer hood.

COUNTRY OF ORIGIN FROM SWEDEN.

Display
A digital display can toggle inbetween l/s, m³/h, °C, hPa. (A PC-setting sets the units to cfm, °F, inHg)
When turning ON by pressing red and blue button the display light is illuminated.

Pressure drop compensation
Use the "back pressure method" and take two measurements, one with a restriction ring and one without, to compensate for restriction and make measurements with high accuracy, even in low pressure ventilation systems. SwemaFlow 126 calculates the uninfluenced flow rate. The backpressure method has proved to have good accuracy, with a total standard measurement error (67 % coverage), including the method, of less than 10 %.

Air Flow scaling
Apart from the Backpressure method it is possible to show the flow multiplied with a Flow Factor. This is useful when balancing the flow, the Flow Factor should be calibrated with the backpressure method. A Hold function can lock the measuring value when it is hard to read the display.

Measuring with SwemaFlow 126
The flow capture’s measuring unit is equipped with two handles which you hold when pressing the capture over the outlet.
The measured values are easily stored by means of a button on the measuring unit. The capture incorporates a rechargable battery and a fast and efficient charger. Charging takes only one to two hours and is sufficient for a normal working day.

Measuring with SwemaFlow 126 Twin
The SwemaFlow 126 Twin has a built-in Bluetooth modem that can communicate with a Swema 3000/md/mdH+, or a SwemaMan 8, if supplemented with a Bluetooth modem. Read more here: SwemaTwin

Barometer, temperature
Densitety compensation is possible since temperature and barometric pressure is measured. Select Real or Standard flow with PC-setting.

Supplied with:

Air flow meter, restriction ring, 300x300 mm exhaust hood, battery charger (230 V), USB cable, manual, carrying case & calibration certificate.


Questions?

Please log in to post a comment
S$ 5400.00
UnitsDiscount
1-
2 ++10%
OPTIONAL:
Technical Data Sheets

Similar Products

Loading Trending..

See More Categories

Electrical
Safety & Construction
Hardware
Warehouse & Transport
Medical Supplies & Equipments
Power Tools
Office & School
F & B Supplies
Chemical Products
O Rings